一、Power requirements
a)voltage:380V±38V;
b)Number of phases:three-phase five-wire system (U, V, W phase line; neutral line; safety ground)
c)Power frequency:50Hz±1Hz;
d)Power supply resistance:less than 0.5Ω.
3.input power:≤35kW。
4.Photography conditions
Tube voltage:40 ~ 125 (Tube current) 16 ~ 200mA
Current time product range:0.5mAs ~ 320mAs。
5working frequency:
Machine operating frequency:≥40kHz。
6.Current time product range:
0.5mAs ~ 320mAs。
7.Overcurrent release characteristics:380V 40A。
(Photography bed)
Bed size:1600 × 730 × 700 (mm) (length × width × height)
Bed movement range:vertical 800mm horizontal 230mm
Bed to film distance:85mm
9、X-ray tube assembly, column
Horizontal movement (along the camera bed): 1620 mm
Move up and down: 1200 mm
Rotate around the center of the arm: ± 180º
10、Maximum size of film holder:356mm × 432mm (14 "× 17")
11、Filter
Stroke: 690mm Gate density: N103 Focal length: 1000mm
二、Features
Compact structure design, excellent coverage performance, support full-body multi-angle photography。
Easy to operate, accurate positioning, optimize the workflow.
X-ray tube components up and down and along the bed movement, photography positioning more convenient
No track, ground track of the integration of bed, the installation is more simple